مرکز تحقیقات چشم

خلاصه مقاله تونومتر دكتر قرايي

امتیاز کاربران

ستاره غیر فعالستاره غیر فعالستاره غیر فعالستاره غیر فعالستاره غیر فعال
 

مرکز تحقیقات چشم>خلاصه مقالات>تونومتر
 

Hamid Gharaei,MD;Abbas Kargozar,MD;Firuze Raygan,MD;Ramin Daneshvar,MD

 

Abstract:

   

This cross-sectional study aimed to investigate if the Tono-Pen, Schiøtz and Perkins tonometers could be used interchangeably in general practice for measuring elevated intraocular pressure. A total of 74 eyes of 37 paediatric patients under general anaesthesia were checked with all 3 tonometers. All of the tonometers gave significantly different measurements from each other. However, with a mean difference of 1.4 mmHg and 95% limits of agreement of –5.7 to +8.6, the greatest agreement was between the Perkins and Tono-Pen tonometers. The Perkins tonometer is a hand-held variant of the Goldmann tonometer (the gold standard for intraocular pressure measures). Therefore the Tono-Pen with its ease of use and safety could be a reliable device for use in general practice.